1.
MEASURE URINE
Urine should not be tested more than 48 hrs. after collection. |
Urine should be at room temp. |
Measure 0.5 ml of centrifuged urine into sealable container. |
|
2.
ADD BUFFER TO URINE
|
Add one drop of Buffer to the container. |
|
Shake 10 seconds. Test within 30 min. |
|
|
|
|
3.
ADD URINE TO WELL
Use a micropipet to transfer 35 µl of buffered urine from step 2 to a
test station. |
|
4.
ADD V-BTA REAGENT TO WELL
Gently shake
V-BTA Reagent for 10 seconds. |
|
Use a micropipet to transfer 35 µl of V-BTA Reagent to the test
station. |
|
|
5.
MIX / WAIT

|
Insufficient mixing is the leading cause of incorrect results. |
|
 |
Mix 10 sec. |
|
Wait 20 sec. |
|
Without splashing, quickly slide test station side-to-side across a flat
surface for 10 seconds to mix fluids. Wait 20 seconds but no longer than 5 minutes. |
|
|
6.
ADD TEST STRIP
|
Lower test strip into well of test station with test pad facing forward. |
|
Read after 30 sec. but no longer than 5 min. |
|